|
16. BREATHING SYSTEMS
An ideal breathing system should be safe and
simple. It should be able to be used for spontaneous and controlled
ventilation. The system would be lightweight, not bulky or complicated and
efficient. It should protect the patient against barotrauma.
Breathing systems include the circle system
(with carbon dioxide reabsorption) and “Mapleson” systems.
Respiratory Physiology
The volume of air inspired during normal
breathing is called the tidal volume (6 to 10 ml/kg). The minute ventilation (MV)
is the tidal volume (TV) times the respiratory rate (RR). The normal adult
minute ventilation is 80 ml/kg/min. Some of the tidal volume air does not enter
the alveoli (where it gives up oxygen and takes up carbon dioxide). It remains
in the oropharynx, trachea and larger airways. This volume of air is called the
anatomical dead space (DS). The normal dead space is about 30% of the tidal
volume. The alveolar ventilation (AV) is the amount of air that is involved in
gas exchange each minute. It is equal to the (TV – DS) x RR.
Expired air contains 5% carbon dioxide and
reduced oxygen (16%). If the patient breathes in his expired air (re-breathing)
he will be breathing high concentrations of carbon dioxide and low
concentrations of oxygen.
Circle System
Circle systems use less gas and volatile
agent, conserve heat and moisture and are suitable for spontaneous ventilation
and intermittent positive pressure ventilation (controlled ventilation or
IPPV).
They can be used a with very low fresh gas
flow (FGF) of less than 1 litre/minute. They must only be used with a very low
fresh gas flow if the anaesthetist can check the inspired oxygen concentration,
there is a carbon dioxide absorber and the inspired oxygen concentration is
greater than 40%.
A circle system is larger, more complex (10
connections) and requires a carbon dioxide absorber.
The circle system consists of seven parts:
the fresh gas flow, inspiratory and expiratory valves, inspiratory and
expiratory tubing, a Y piece connector, reservoir bag, overflow or airway pressure
limiting (APL) valve and the carbon dioxide absorbent container. There are
several different ways of arranging the parts. To prevent rebreathing, the
fresh gas flow must not enter between the expiratory valve and the patient, the
overflow valve must not be located between the patient and the inspiratory
valve, and the inspiratory and expiratory valves must be located between the
patient and the reservoir bag on both the inspiratory and expiratory limbs of
the circuit.
The fresh gas flow enters the inspiratory
limb of the circle and passes though the inspiratory valve to the patient.
Exhaled gas passes along the expiratory limb though the expiratory valve to a
carbon dioxide absorber and back to the patient.
There are several common carbon dioxide absorbents
(e.g. soda lime). In general, they contain a hydroxide that reacts with carbon
dioxide. Heat and water are produced as by-products. They contain a chemical
indicator which changes colour when the soda lime is exhausted. The
anaesthetist must know which chemical indicator is used. Different chemical
indicators change to different colours.
A circle system can be used without soda lime
but re-breathing and carbon dioxide retention can occur. The risk of
re-breathing depends on the arrangement of the parts, the fresh gas flow and
the ventilation. To prevent re-breathing the fresh gas flow should be 60
ml/kg/min and ventilate at thee times normal minute ventilation, or set the
fresh gas flow to alveolar ventilation and ventilate at thee times normal minute
ventilation.
Vaporisers can be placed in their usual
position on the back bar (vaporiser out of circuit VOC) or can rarely be placed
in the circle breathing system (vaporiser in circuit VIC). Vaporisers made to
work with compressed gas (plenum) or drawover vaporisers must never be placed
in the circuit. Gas expired from the patient will contain some volatile
anaesthetic agent. If this is allowed to recirculate though the vaporiser it
will continue to increase the volatile concentration above the concentration
which has been selected on the vaporiser. Vaporisers should only be placed in
circuit if they are made to be used in a circle breathing system and agent
concentration monitoring is available.
Trichloroethylene must not be used with
carbon dioxide absorbers due to production of toxic products.
Mapleson Breathing Systems
The Mapleson breathing systems have no valves
to direct gases to and from the patient. There is no carbon dioxide absorber.
The fresh gas flow must wash out the expired carbon dioxide in the breathing
system. The parts of a Mapleson breathing system are a reservoir bag, tubing,
fresh gas flow, APL valve and patient connector. The Mapleson breathing systems
are simple and inexpensive. They require high fresh gas flow to prevent re-breathing
and the fresh gas flow rate may need to be altered when changing from
spontaneous to controlled ventilation. They do not conserve heat or moisture.
The Mapleson A, B and C breathing systems have the APL valve close to the
patient where it may be difficult to access. The Mapleson E and F breathing
systems are difficult to scavenge. If there is a fall in fresh gas flow with
the Mapleson breathing systems there is a risk of re-breathing.
There are different ways of arranging the
parts.
The Mapleson A (Magill) breathing system is
efficient for spontaneous ventilation. Fresh gas flow should equal minute
ventilation. It is inefficient for controlled ventilation. Fresh gas flow must
be 2 to 3 times minute ventilation to prevent re-breathing.
The Mapleson B and C breathing systems are
rarely used for anaesthesia. They are used for resuscitation. Fresh gas flow
for controlled ventilation should be 2 to 2.5 times minute ventilation.
The Mapleson D breathing system is
inefficient for spontaneous ventilation. The flow rate should be 150 to 250
ml/kg/min. It is efficient for controlled ventilation. Fresh gas flow should be
70 ml/kg/min.
The Mapleson E (Ayres T piece) breathing
system is used in children because it has a very low resistance and minimal
dead space. The reservoir limb should be larger than the tidal volume and fresh
gas flow should be 2 to 3 times minute ventilation.
The Mapleson F (Jackson Rees modification of
the Ayres T piece) breathing system is a Mapleson E breathing system with an
open bag attached to the expiratory limb. The bag allows easy controlled
ventilation and visual assessment of spontaneous ventilation. Fresh gas flow
should be 2 to 3 times minute ventilation.
|